10 CFR §35.61
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)A licensee shall calibrate the survey instruments used to show compliance with this part and 10 CFR Part 20 before first use, annually, and following a repair that affects the calibration. A licensee shall—
- (b)A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
- (c)A licensee shall retain a record of each survey instrument calibration in accordance with § 35.2061.