10 CFR §35.647
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)A licensee providing mobile remote afterloader service shall—
- (b)In addition to the periodic spot-checks required by § 35.643, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of—
- (1)Electrical interlocks on treatment area access points;
- (2)Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;
- (3)Viewing and intercom systems;
- (4)Applicators, source transfer tubes, and transfer tube-applicator interfaces;
- (5)Radiation monitors used to indicate room exposures;
- (6)Source positioning (accuracy); and
- (7)Radiation monitors used to indicate whether the source has returned to a safe shielded position.
- (c)In addition to the requirements for checks in paragraph (b) of this section, a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
- (d)If the results of the checks required in paragraph (b) of this section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
- (e)A licensee shall retain a record of each check required by paragraph (b) of this section in accordance with § 35.2647.