21 CFR §820.200
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Where servicing is a specified requirement, each manufacturer shall establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements.
- (b)Each manufacturer shall analyze service reports with appropriate statistical methodology in accordance with § 820.100.
- (c)Each manufacturer who receives a service report that represents an event which must be reported to FDA under part 803 of this chapter shall automatically consider the report a complaint and shall process it in accordance with the requirements of § 820.198.
- (d)Service reports shall be documented and shall include: