21 CFR §862.1350
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A continuous glucose monitor (CGM) secondary alarm system is identified as a device intended to be used as a secondary alarm for a CGM to enable immediate awareness for potential clinical intervention to help assure patient safety.
- (b)Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special controls for this device are:
- (1)Devices being marketed must include appropriate measures to protect against unauthorized access to data and unauthorized modification of data.
- (2)The labeling must prominently and conspicuously display a warning that states “Dosing decisions should not be made based on this device. The user should follow instructions on the continuous glucose monitoring system.”
- (3)The labeling for the device must include a statement that reads “This device is not intended to replace self-monitoring practices as advised by a physician.”