21 CFR §868.1505
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A ventilatory electrical impedance tomograph is a prescription non-invasive, non-radiological ventilatory device that provides an assessment of local impedance variation within a cross-section of a patient's thorax.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
- (i)Characterization of device parameters, including signal-to-noise ratio, voltage accuracy, drift, reciprocity accuracy, amplitude response, position error, and ringing;
- (ii)Real time evaluation of local impedance variation;
- (iii)Plethysmogram accuracy testing; and
- (iv)Use life testing of reusable components.
- (3)Performance data must validate reprocessing instructions for any reusable components of the device.
- (4)Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device.
- (5)Software verification, validation, and hazard analysis must be performed.
- (6)Labeling must include the following:
- (i)Guidance for interpretation of the images generated;
- (ii)A warning that the device should be removed before use of a defibrillator, or defibrillator interaction information based on defibrillator performance testing with the device;
- (iii)A use life for any reusable components; and
- (iv)Instructions for reprocessing any reusable components.