- (a)Identification. A high flow humidified oxygen delivery device is a prescription device that delivers high flow oxygen with humidification for patients who are suffering from respiratory distress and/or hypoxemia.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, including the following:
- (3)Performance data must validate cleaning instructions for any reusable components of the device.
- (4)Electrical safety, thermal safety, mechanical safety, electromagnetic compatibility, and radiofrequency identification testing must be performed.
- (5)Software verification, validation, and hazard analysis must be performed.
- (6)Labeling must include:
- (i)A description of available FiO2 ranges for different flowrates and inlet gas pressures;
- (ii)Instructions for applicable flowrates for all intended populations;
- (iii)A warning that patients on high flow oxygen are acute and require appropriate monitoring, to include pulse oximetry;
- (iv)A warning regarding the risk of condensation at low set temperatures and certain flows; and
- (v)A description of all alarms and their functions.