21 CFR §870.1420
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A coronary artery disease risk indicator using acoustic heart signals is a device that records heart sounds including murmurs and vibrations to calculate a patient-specific risk of presence of coronary artery disease, as an aid in cardiac analysis and diagnosis.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)Clinical performance testing must fulfill the following:
- (2)Acoustic performance testing must evaluate microphone sensitivity, sound acquisition bandwidth, and amplitude accuracy. The acoustic sensor specifications and mechanism used to capture heart sounds must be described.
- (3)A scientific justification for the validity of the algorithm(s) must be provided. This justification must fulfill the following:
- (i)All inputs and outputs of the algorithm must be fully described;
- (ii)The procedure for segmenting, characterizing, and classifying the acoustic signal must be fully described; and
- (iii)This justification must include verification of the algorithm calculations and validation using an independent data set.
- (4)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (5)Software verification, validation, and hazard analysis must be performed.
- (6)Human factors/usability testing must demonstrate that the user can correctly use the device, including device placement, based solely on reading the directions for use.
- (7)Performance data must demonstrate the electromagnetic compatibility and electrical safety of the device.
- (8)Labeling must include the following:
- (i)A description of what the device measures and outputs to the user;
- (ii)Instructions for proper placement of the device;
- (iii)Instructions on care and cleaning of the device;
- (iv)Warnings identifying sensor acquisition factors that may impact measurement results and instructions for mitigating these factors; and
- (v)The expected performance of the device for all intended use populations and environments.