21 CFR §870.2790
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A photoplethysmograph analysis software device for over-the-counter use analyzes photoplethysmograph data and provides information for identifying irregular heart rhythms. This device is not intended to provide a diagnosis.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)Clinical performance testing must demonstrate the performance characteristics of the detection algorithm under anticipated conditions of use.
- (2)Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the detection algorithm and its inputs and outputs.
- (3)Non-clinical performance testing must demonstrate the ability of the device to detect adequate photoplethysmograph signal quality.
- (4)Human factors and usability testing must demonstrate the following:
- (5)Labeling must include:
- (i)Hardware platform and operating system requirements;
- (ii)Situations in which the device may not operate at an expected performance level;
- (iii)A summary of the clinical performance testing conducted with the device;
- (iv)A description of what the device measures and outputs to the user; and
- (v)Guidance on interpretation of any results.