21 CFR §874.4450
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A powered insertion system for a cochlear implant electrode array is a prescription device used to assist in placing an electrode array into the cochlea.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including evaluation of all adverse events.
- (2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
- (i)Verification of cochlear implant attachment force, release force, and insertion speed;
- (ii)Testing to demonstrate the device does not damage or degrade the cochlear implant (including the lead and array portions of the cochlear implant); and
- (iii)Comparison testing with manual insertion to evaluate:
- (3)Usability testing in a simulated hospital environment with an anatomically relevant model (e.g., cadaver testing) that evaluates the following:
- (4)Changes in cochlear implant compatibility are determined to significantly affect the safety or effectiveness of the device and must be validated through performance testing or a rationale for omission of any testing.
- (5)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (6)Performance testing must demonstrate the electromagnetic compatibility, electrical safety, and thermal safety of the device.
- (7)The patient-contacting components of the device must be demonstrated to be sterile and non-pyrogenic.
- (8)Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
- (9)Software verification, validation, and hazard analysis must be performed for any software components of the device.
- (10)Labeling must include: