21 CFR §876.5011
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. A metallic biliary stent system for benign strictures is a prescription device intended for the treatment of benign biliary strictures. The biliary stents are intended to be left indwelling for a limited amount of time and subsequently removed. The device consists of a metallic stent and a delivery system intended to place the biliary stent in the bile duct. This device type is not intended for use in the vasculature.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)Clinical performance testing must demonstrate or provide the following:
- (i)The ability to safely place and subsequently remove the stent after the maximum labeled indwell period.
- (ii)All adverse event data including bile duct obstruction and trauma to the bile duct.
- (iii)The stent resolves strictures during the maximum labeled indwell period.
- (iv)Stricture resolution is maintained post-stent removal.
- (2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
- (i)Corrosion testing to demonstrate that the stent maintains its integrity during indwell and does not release potentially toxic levels of leachables.
- (ii)Stent dimensional testing supports the intended use.
- (iii)Compression and expansion forces must be characterized.
- (iv)The delivery catheter must deliver the stent to the intended location and the stent must not be adversely impacted by the delivery catheter during deployment and catheter withdrawal.
- (v)The delivery system must withstand clinically anticipated forces.
- (vi)Compatibility in a magnetic resonance environment.
- (3)All patient contacting components of the device must be demonstrated to be biocompatible.
- (4)Performance data must demonstrate the sterility of the device components intended to be provided sterile.
- (5)Shelf life testing must demonstrate that the device maintains its performance characteristics and that packaging maintains sterility for the duration of the labeled shelf life.
- (6)Labeling for the device must include:
- (i)A detailed summary of the clinical testing including device effectiveness, and device- and procedure-related adverse events.
- (ii)Appropriate warning(s) to accurately ensure usage of the device for the intended patient population.
- (iii)Shelf life.
- (iv)Compatibility information for use in the magnetic resonance environment.
- (v)Stent foreshortening information supported by dimensional testing.
- (1)Clinical performance testing must demonstrate or provide the following: