- (a)Identification. This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.
- (2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:
- (i)Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract;
- (ii)Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and
- (iii)Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.
- (3)Clinical performance testing must demonstrate the device performs as intended and evaluate the following:
- (4)Physician and patient device labeling must state:
- (i)The clinical benefit of the device as assessed by using percent total body weight loss;
- (ii)Treatment must be offered in combination with diet and exercise;
- (iii)Instructions on how to use the device as intended including how to avoid interaction with medication; and
- (iv)The shelf life of the device.