- (a)Identification. An autofluorescence detection device for general surgery and dermatological use is an adjunct tool that uses autofluorescence to detect tissues or structures. This device is not intended to provide a diagnosis.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)In vivo testing under anticipated conditions of use must characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.
- (2)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (3)Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical, and thermal safety of the device.
- (4)Software verification, validation, and hazard analysis must be performed.
- (5)Performance testing must demonstrate the sterility of patient-contacting components of the device.
- (6)Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
- (7)Performance testing must demonstrate laser and light safety for eye, tissue, and skin.
- (8)Labeling must include the following: