21 CFR §878.4685
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Identification. An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
- (2)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (3)Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
- (4)Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
- (5)Software verification, validation, and hazard analysis must be performed.
- (6)Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
- (7)Physician labeling must include:
- (i)Information on how the device operates and the typical course of treatment;
- (ii)A detailed summary of the device's technical parameters;
- (iii)Validated methods and instructions for reprocessing of any reusable components; and
- (iv)Instructions for preventing hearing loss by use of hearing protection.
- (8)Patient labeling must include:
- (i)Relevant contraindications, warnings, precautions, adverse effects, and complications;
- (ii)Information on how the device operates and the typical course of treatment;
- (iii)The probable risks and benefits associated with the use of the device;
- (iv)Post-procedure care instructions; and
- (v)Alternative treatments.