- (a)Identification. The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized.
- (2)The cleaning and disinfection instructions for the device must be validated.
- (3)The device must be demonstrated to be biocompatible.
- (4)Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device.
- (5)Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device.
- (6)Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen.
- (7)Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance.
- (8)Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.