- (a)Identification. A vibratory counter-stimulation device is a prescription device that provides electrically powered mechanical vibration to improve the quality of sleep in patients with primary Restless Legs Syndrome.
- (b)Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9. The special controls for this device are:
- (1)Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
- (2)If the device contains software or firmware, appropriate verification, validation, and hazard analysis must be performed.
- (3)The elements of the device that contact the patient must be assessed to be biocompatible.
- (4)Non-clinical testing data (including vibration frequency, amplitude, and acceleration) must demonstrate that the device performs as intended under anticipated conditions of use.
- (5)Labeling must include:
- (i)Specific information pertinent to use of the device by the intended patient population and the treatment regimen;
- (ii)Warning to only use the device on normal, intact, clean, healthy skin;
- (iii)Warning to not use the device if the user has leg skin disorders, such as eczema, psoriasis, cellulitis, non-healing wounds;
- (iv)Warning to discontinue use if Restless Leg Syndrome symptoms worsen; and
- (v)Instructions for end users to contact the device manufacturer and MedWatch in case they experience any adverse events when using this device.