- (a)Identification. A fetal head elevator is a prescription device consisting of a mechanism that elevates the fetal head to facilitate delivery during a Caesarean section.
- (b)Classification. Class II (special controls). The special controls for this device are:
- (1)The patient-contacting components of the device must be demonstrated to be biocompatible.
- (2)Performance data must demonstrate the sterility of patient-contacting components of the device.
- (3)Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
- (4)Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- (i)Reliability testing of device deployment and retrieval under relevant use conditions must be conducted.
- (ii)Testing of the maximum force applied to the fetal head in an anatomic model must be conducted.
- (iii)Testing of uniform application of the elevator mechanism on the fetal head must be conducted.
- (5)Labeling must include the following: