MedRadio transmitters operate in the 401-406 MHz, 413-419 MHz, 426-432 MHz, 438-444 MHz, 451-457 MHz, and 2360-2400 MHz bands. The FCC does not specify a channeling scheme for MedRadio systems.
  1. (a)MedRadio transmitters associated with medical implant devices, which incorporate a frequency monitoring system as set forth in § 95.2559(a), may transmit on any frequency in the 401-406 MHz band.
  2. (b)MedRadio transmitters associated with medical implant devices, which do not incorporate a frequency monitoring system as set forth in § 95.2559(a), may transmit on any frequency in the 401-402 MHz or 405-406 MHz bands, or on the frequency 403.65 MHz in the 402-405 MHz band.
  3. (c)MedRadio transmitters associated with medical body-worn devices, regardless of whether a frequency monitoring system as set forth in § 95.2559(a) is employed, may transmit on any frequency in the 401-402 MHz or 405-406 MHz bands.
  4. (d)MedRadio transmitters that are used externally to evaluate the efficacy of a more permanent medical implant device, regardless of whether a frequency monitoring system as set forth in § 95.2559(a) is employed, may operate on any frequency in the 402-405 MHz band, provided that:
    1. (1)Such external body-worn operation is limited solely to evaluating with a patient the efficacy of a fully implanted permanent medical device that is intended to replace the temporary body-worn device;
    2. (2)RF transmissions from the external device must cease following the patient evaluation period, which may not exceed 30 days, except where a health care practitioner determines that additional time is necessary due to unforeseen circumstances;
    3. (3)The maximum output power of the temporary body-worn device must not exceed 200 nW EIRP; and
    4. (4)The temporary body-worn device must comply fully with all other MedRadio rules applicable to medical implant device operation in the 402-405 MHz band.
  5. (e)Only MedRadio transmitters that are part of a Medical Micropower Network (MMN) may operate in the 413-419 MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz bands. Each MedRadio transmitter that is part of an MMN must be capable of operating in each of the following bands: 413-419 MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz. All MedRadio transmitters that are part of a single MMN must operate in the same band.
  6. (f)Only MedRadio transmitters that are part of a Medical Body Area Network (MBAN) may operate in the 2360-2400 MHz band.