Each MedRadio transmitter type must be designed such that the MedRadio equivalent isotropically radiated power (M-EIRP) does not exceed the limits in this section. Compliance with these limits must be determined as set forth in § 95.2569.
- (a)Transmitters subject to frequency monitoring—401-406 MHz. For MedRadio transmitters that are not excepted under § 95.2559(b) from the frequency monitoring requirements of § 95.2559(a):
- (b)Transmitters excepted from frequency monitoring—401-402 MHz and 405-406 MHz. For MedRadio transmitters that are excepted under § 95.2559(b)(2) or (3) from the frequency monitoring requirements of § 95.2559(a):
- (c)Transmitters excepted from frequency monitoring—403.65 MHz. For MedRadio transmitters that are excepted under § 95.2559(b)(4) from the frequency monitoring requirements of § 95.2559(a), the M-EIRP must not exceed 100 nanowatts in the 300 kHz bandwidth centered at 403.65 MHz.
- (d)Transmitters—other frequency bands. For MedRadio transmitters operating in the 413-419 MHz, 426-432 MHz, 438-444 MHz, or 451-457 MHz bands:
- (e)Transmitters—2360-2390 MHz band. For MedRadio transmitters operating in the 2360-2390 MHz band, the M-EIRP over the bands of operation must not exceed the lesser of zero dBm (1 mW) or 10 log (B) dBm, where B is the MedRadio 20 dB emission bandwidth in megahertz.
- (f)Transmitters—2390-2400 MHz band. For MedRadio transmitters operating in the 2390-2400 MHz band, the M-EIRP over the bands of operation must not exceed the lesser of 13 dBm (20 mW) or 16 + 10 log (B) dBm, where B is the MedRadio 20 dB emission bandwidth in megahertz.