40 CFR §1065.267
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Application. You may use a gas chromatograph with a flame ionization detector (GC-FID) to measure CH4 and C2H6 concentrations of diluted exhaust for batch sampling. While you may also use a nonmethane cutter to measure CH4, as described in § 1065.265, use a reference procedure based on a gas chromatograph for comparison with any proposed alternate measurement procedure under § 1065.10.
- (b)Component requirements. We recommend that you use a GC-FID that meets the specifications in § 1065.205 and that the measurement be done according to SAE J1151 (incorporated by reference, see § 1065.1010). The GC-FID must meet the linearity verification in § 1065.307.