- (a)Scope and frequency. If you measure CH4, C2H6, NMHC, or NMNEHC using an FTIR analyzer, verify the amount of interference after initial analyzer installation and after major maintenance.
- (b)Measurement principles. Certain species can interfere with analyzers by causing a response similar to the target analyte. If the analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, a correct result depends on simultaneously conducting these other measurements to test the compensation algorithms during the analyzer interference verification.
- (c)System requirements. An FTIR analyzer must have combined interference that is within ±2% of the flow-weighted mean concentration of CH4, NMHC, or NMNEHC expected at the standard, though we strongly recommend a lower interference that is within ±1%.
- (d)Procedure. Perform the interference verification for an FTIR analyzer using the same procedure that applies for N2O analyzers in § 1065.375(d).