40 CFR §63.8688
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)You must install, operate, and maintain each continuous parameter monitoring system (CPMS) according to the following:
- (1)The CPMS must complete a minimum of one cycle of operation for each successive 15-minute period.
- (2)To determine the 3-hour average, you must:
- (i)Have a minimum of four successive cycles of operation to have a valid hour of data.
- (ii)Have valid data from at least three of four equally spaced data values for that hour from a CPMS that is not out-of-control according to your site-specific monitoring plan.
- (iii)Determine the 3-hour average of all recorded readings for each operating day, except as stated in § 63.8690(c). You must have at least two of the three hourly averages for that period using only hourly average values that are based on valid data (i.e., not from out-of-control periods).
- (3)You must record the results of each inspection, calibration, and validation check.
- (b)For each temperature monitoring device, you must meet the requirements in paragraph (a) of this section and the following:
- (1)Locate the temperature sensor in a position that provides a representative temperature.
- (2)For a noncryogenic temperature range, use a temperature sensor with a minimum measurement sensitivity of 2.8 °C or 1.0 percent of the temperature value, whichever is larger.
- (3)If a chart recorder is used, it must have a sensitivity in the minor division of at least 20 °F.
- (4)Perform an accuracy check at least semiannually or following an operating parameter deviation:
- (i)According to the procedures in the manufacturer's documentation; or
- (ii)By comparing the sensor output to redundant sensor output; or
- (iii)By comparing the sensor output to the output from a calibrated temperature measurement device; or
- (iv)By comparing the sensor output to the output from a temperature simulator.
- (5)Conduct accuracy checks any time the sensor exceeds the manufacturer's specified maximum operating temperature range or install a new temperature sensor.
- (6)At least quarterly or following an operating parameter deviation, perform visual inspections of components if redundant sensors are not used.
- (c)For each pressure measurement device, you must meet the requirements of paragraph (a) of this section and the following:
- (1)Locate the pressure sensor(s) in, or as close as possible, to a position that provides a representative measurement of the pressure.
- (2)Use a gauge with a minimum measurement sensitivity of 0.12 kiloPascals or a transducer with a minimum measurement sensitivity of 5 percent of the pressure range.
- (3)Check pressure tap pluggage daily. Perform an accuracy check at least quarterly or following an operating parameter deviation:
- (4)Conduct calibration checks any time the sensor exceeds the manufacturer's specified maximum operating pressure range or install a new pressure sensor.
- (5)At least monthly or following an operating parameter deviation, perform a leak check of all components for integrity, all electrical connections for continuity, and all mechanical connections for leakage.
- (6)At least quarterly or following an operating parameter deviation, perform visible inspections on all components if redundant sensors are not used.
- (d)For monitoring parameters other than temperature and pressure drop, you must install and operate a CPMS to provide representative measurements of the monitored parameters.
- (e)For each flare, you must install a device (including but not limited to a thermocouple, an ultraviolet beam sensor, or an infrared sensor) capable of continuously detecting the presence of a pilot flame.
- (f)As an option to installing the CPMS specified in paragraph (a) of this section, you may install a continuous emissions monitoring system (CEMS) or a continuous opacity monitoring system (COMS) that meets the applicable requirements in § 63.8 according to Table 7 to this subpart and the applicable performance specifications of 40 CFR part 60, appendix B.
- (g)For each monitoring system required in this section, you must develop and make available for inspection by the permitting authority, upon request, a site-specific monitoring plan that addresses the following:
- (1)Installation of the CPMS, CEMS, or COMS sampling probe or other interface at a measurement location relative to each affected process unit such that the measurement is representative of control of the exhaust emissions (e.g., on or downstream of the last control device);
- (2)Performance and equipment specifications for the sample interface, the pollutant concentration or parametric signal analyzer, and the data collection and reduction system; and
- (3)Performance evaluation procedures and acceptance criteria (e.g., calibrations).
- (h)In your site-specific monitoring plan, you must also address the following:
- (1)Ongoing operation and maintenance procedures in accordance with the general requirements of § 63.8(c)(1)(ii), (c)(3), (c)(4)(ii), and (c)(7) and (8);
- (2)Ongoing data quality assurance procedures in accordance with the general requirements of § 63.8(d); and
- (3)Ongoing recordkeeping and reporting procedures in accordance with §§ 63.8693 and 63.8694 and the general requirements of § 63.10(e)(1) and (e)(2)(i).
- (i)You must conduct a performance evaluation of each CPMS, CEMS, or COMS in accordance with your site-specific monitoring plan.
- (j)You must operate and maintain the CPMS, CEMS, or COMS in continuous operation according to the site-specific monitoring plan.