42 CFR §493.1236

Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov
  1. (a)The laboratory must review and evaluate the results obtained on proficiency testing performed as specified in subpart H of this part.
  2. (b)The laboratory must verify the accuracy of the following:
    1. (1)Any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS-approved proficiency testing program.
    2. (2)Any analyte, specialty or subspecialty assigned a proficiency testing score that does not reflect laboratory test performance (that is, when the proficiency testing program does not obtain the agreement required for scoring as specified in subpart I of this part, or the laboratory receives a zero score for nonparticipation, or late return of results).
  3. (c)At least twice annually, the laboratory must verify the accuracy of the following:
    1. (1)Any test or procedure it performs that is not included in subpart I of this part.
    2. (2)Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.
  4. (d)All proficiency testing evaluation and verification activities must be documented.