42 CFR §493.1267
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
For blood gas analyses, the laboratory must perform the following:
- (a)Calibrate or verify calibration according to the manufacturer's specifications and with at least the frequency recommended by the manufacturer.
- (b)Test one sample of control material each 8 hours of testing using a combination of control materials that include both low and high values on each day of testing.
- (c)Test one sample of control material each time specimens are tested unless automated instrumentation internally verifies calibration at least every 30 minutes.
- (d)Document all control procedures performed, as specified in this section.