42 CFR §84.172

Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov
  1. (a)Resistance to airflow will be measured in the tight-fitting respiratory inlet covering of a complete particulate respirator mounted on a test fixture with air flowing at continuous rate of 85 ±2 liters per minute, before each test conducted in accordance with § 84.173.
  2. (b)Resistance of a complete tight-fitting powered air-purifying particulate respirator system will be measured with the blower off if the manufacturer indicates that the respirator should not be doffed in the event of a blower failure.
  3. (c)The maximum allowable resistance requirements for air-purifying particulate respirators are as follows: