42 CFR §84.181
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Powered air-purifying respirators must be designed to allow for proper communication while worn.
- (b)A Modified Rhyme Test 7 will be used to test the wearer's ability to communicate efficiently.
- (c)The communications requirement is met if the overall performance rating is greater than or equal to 70 percent.