42 CFR §84.91
Verified against eCFR.gov as of June 20, 2026View official text on eCFR.gov ↗
- (a)Resistance to exhalation airflow will be measured in the facepiece or mouthpiece of open-circuit apparatus with air flowing at a continuous rate of 85 liters per minute.
- (b)The exhalation resistance of demand apparatus shall not exceed 25 mm. (1 inch) water-column height.
- (c)The exhalation resistance of pressure-demand apparatus shall not exceed the static pressure in the facepiece by more than 51 mm. (2 inches) water-column height.
- (d)The static pressure (at zero flow) in the facepiece shall not exceed 38 mm. (1.5 inches) water-column height.
- (e)Resistance to exhalation airflow will be measured in the facepiece or mouthpiece of closed-circuit apparatus with a breathing machine as described in § 84.88, and the exhalation resistance shall not exceed 51 mm. (2 inches) water-column height.